HD300 V2.0 11/07
11
Specifications
General Specifications
Display Dual Display Multi-function LCD with 9999 counts
Measurements Air Velocity: m/s, km/h, ft/min, knots, mph;
Air Flow: CMM (m
3
/min) and CFM (ft
3
/min);
Air Temperature (through vane) and Surface Temperature
(using the IR thermometer function): °C and °F
Data Hold Freezes displayed reading
Sampling rate 1 reading per second
Sensors Air velocity/Air flow sensor: Conventional angled vane arms
with low-friction ball bearing. Air temperature sensor: Precision
thermistor (built into the vane assembly); Surface temperature
via non-contact IR sensor.
IR Distance to Spot ratio 30:1
IR Spectral response 6 to 14µm
IR Emissivity 0.95 fixed
MIN-MAX-AVG Record and Recall lowest, highest, and average readings
Auto Power OFF Automatic shut off after 15 minutes (can be disengaged)
PC Interface USB PC Communication with supplied software and cable for
data acquisition
Over range indication Dashes appears on the LCD
Low battery indication Battery symbol appears on the LCD
Power supply 9V Battery
Operating conditions Meter: 32 to 122
o
F (0 to 50
o
C); 80% RH max.
Sensor: 32 to 140
o
F (0 to 60
o
C)
Dimensions / Weight Main instrument: 8.0 x 3.0 x1.9" (203 x 75 x 50mm)
Sensor head: 2.8” (72mm) diameter
Weight 9.8 oz. (280g)
Commenti su questo manuale