Pagina 2 - Introduction

MN16-EU V1.4 03/12 10Specifications Function Range Resolution Accuracy 4V 0.001V 40V 0.01V 400V 0.1V ±(1.2% reading + 2 digits) DC Voltage (V DC)

Pagina 3 - Symbols and Annunciators

MN16-EU V1.4 03/12 11Function Range Resolution Accuracy 5.000Hz 0.001Hz 50.00Hz 0.01Hz ±(1.5% reading + 5 digits) 500.0Hz 0.1Hz 5.000kHz 0.001kHz

Pagina 4 - Operating Instructions

MN16-EU V1.4 03/12 12PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is equipment for con

Pagina 5

MN16-EU V1.4 03/12 2Introduction Congratulations on your purchase of the Extech MN16 Autoranging Multimeter. This meter measures AC/DC Voltage, AC

Pagina 6

MN16-EU V1.4 03/12 3 Input Protection Limits Function Maximum Input V DC or V AC 600VDC/AC mA AC/DC 400mA AC/DC A AC/DC 10A AC/DC (20A for 30

Pagina 7

MN16-EU V1.4 03/12 4Operating Instructions WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be

Pagina 8

MN16-EU V1.4 03/12 5DC CURRENT MEASUREMENTS CAUTION: Do not make 20A current measurements for longer than 30 seconds. Exceeding 30 seconds may ca

Pagina 9 - Maintenance

MN16-EU V1.4 03/12 6RESISTANCE MEASUREMENTS WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors

Pagina 10 - Specifications

MN16-EU V1.4 03/12 7TEMPERATURE MEASUREMENTS 1. Set the function switch to the ºF or ºC position. 2. Insert the Temperature Probe into the input